Neurothrombectomy Devices for the Treatment of Acute Ischemic Stroke: State of the Evidence FREE

William L. Baker, PharmD; Jennifer A. Colby, PharmD; Vanita Tongbram, MBBS, MPH; Ripple Talati, PharmD; Isaac E. Silverman, MD; C. Michael White, PharmD; Jeffrey Kluger, MD; and Craig I. Coleman, PharmD
[+] Article and Author Information

From the University of Connecticut/Hartford Hospital Evidence-based Practice Center, Hartford, Connecticut.


Disclaimer: The authors of this report are responsible for its content. Statements in the report should not be construed as endorsement by the Agency for Healthcare Research and Quality or the U.S. Department of Health and Human Services.

Grant Support: This article is based on research conducted by the University of Connecticut/Hartford Hospital Evidence-based Practice Center under contract 290-2007-10067-I with the Agency for Healthcare Research and Quality.

Potential Conflicts of Interest: Drs. Baker, Colby, Tongbram, Talati, White, and Coleman: Grant (money to institution): Agency for Healthcare Research and Quality. Dr. Silverman: Payment for writing or reviewing the manuscript: Agency for Healthcare Research and Quality. Dr. Kluger: Grant (money to institution): Agency for Healthcare Research and Quality, Boston Scientific, Astellas; Consultancy: Sanofi-Aventis; Payment for lectures including service on speakers' bureaus: Medtronic, Sanofi-Aventis. Disclosures can also be viewed at www.acponline.org/authors/icmje/ConflictOfInterestForms.do?msNum=M10-2218.

Requests for Single Reprints: Craig I. Coleman, PharmD, University of Connecticut/Hartford Hospital Evidence-based Practice Center, 80 Seymour Street, Hartford, CT 06102-5037; e-mail, ccolema@harthosp.org.

Current Author Addresses: Dr. Baker: University of Connecticut School of Pharmacy & Medicine, 263 Farmington Avenue, Farmington, CT 06030.

Drs. Colby, Tongbram, Talati, Silverman, White, Kluger, and Coleman: University of Connecticut/Hartford Hospital Evidence-based Practice Center, 80 Seymour Street, Hartford, CT 06102-5037.

Author Contributions: Conception and design: W.L. Baker, I.E. Silverman, C.M. White, J. Kluger, C.I. Coleman.

Analysis and interpretation of the data: W.L. Baker, J.A. Colby, V. Tongbram, R. Talati, I.E. Silverman, C.M. White, C.I. Coleman.

Drafting of the article: W.L. Baker, J.A. Colby, V. Tongbram, R. Talati, I.E. Silverman, C.I. Coleman.

Critical revision of the article for important intellectual content: W.L. Baker, I.E. Silverman, C.M. White, C.I. Coleman.

Final approval of the article: W.L. Baker, J.A. Colby, V. Tongbram, R. Talati, I.E. Silverman, C.M. White, J. Kluger, C.I. Coleman.

Provision of study materials or patients: C.I. Coleman.

Statistical expertise: W.L. Baker, J.A. Colby, V. Tongbram, C.I. Coleman.

Obtaining of funding: C.M. White, C.I. Coleman.

Administrative, technical, or logistic support: J.A. Colby, V. Tongbram, I.E. Silverman, C.M. White, J. Kluger, C.I. Coleman.

Collection and assembly of data: W.L. Baker, J.A. Colby, V. Tongbram, R. Talati, C.I. Coleman.


Ann Intern Med. 2011;154(4):243-252. doi:10.7326/0003-4819-154-4-201102150-00306
Text Size: A A A

Background: Acute ischemic strokes are associated with poor outcomes and high health care burden. Evidence exists evaluating the use of neurothrombectomy devices in patients receiving currently recommended treatments that may have limited efficacy.

Purpose: To describe the state of the evidence supporting use of neurothrombectomy devices in the treatment of acute ischemic stroke.

Data Sources: MEDLINE, SCOPUS, the Cochrane Central Register of Controlled Trials, the Cochrane Database of Systematic Reviews, and Web of Science were searched, without language restrictions, from their inception through May 2010. The MEDLINE and Cochrane Central Register of Controlled Trials searches were updated through November 2010.

Study Selection: Two independent investigators screened citations for human studies of any design or case series or case reports of patients with an acute ischemic stroke that evaluated a neurothrombectomy device and reported at least 1 clinical effectiveness outcome or harm.

Data Extraction: Using standardized protocols, 2 independent investigators extracted information about study characteristics and outcomes, and a third reviewer resolved disagreement.

Data Synthesis: 87 articles met eligibility criteria, including 18 prospective single-group studies, 7 noncomparative retrospective studies, and 62 case series or case reports. Two U.S. Food and Drug Administration (FDA)–cleared devices, the MERCI Retriever (Concentric Medical, Mountain View, California) (40%) and the Penumbra System (Penumbra, Alameda, California) (9%), represented a large portion of the available data. All prospective and retrospective studies provided data on successful recanalization with widely varying rates (43% to 78% with the MERCI Retriever and 83% to 100% with the Penumbra System). Rates of harms, including symptomatic (16 studies; 0% to 10% with the MERCI Retriever and 0% to 11% with the Penumbra System) or asymptomatic (13 studies; 28% to 43% and 1% to 30%, respectively) intracranial hemorrhage and vessel perforation or dissection (11 studies; 0% to 7% and 0% to 5%, respectively), also varied by device. Predictors of harm included older age, history of stroke, and higher baseline stroke severity scores, whereas successful recanalization was the sole predictor of good outcomes.

Limitations: Most available data are from single-group, noncomparative studies. In addition, the patient population most likely to benefit from these devices is undetermined.

Conclusion: Currently available neurothrombectomy devices offer intriguing treatment options in patients with acute ischemic stroke. Future trials should use a randomized design, with adequate power to show equivalency or noninferiority between competing strategies or devices, and strive to identify populations that are most likely to benefit from use of neurothrombectomy devices.

Primary Funding Source: Agency for Healthcare Research and Quality.

Stroke is the third leading cause of death in the United States—an estimated 795 000 new or recurrent events occur annually, and most are classified as ischemic in nature (12). Patient variables associated with worse outcomes and increased health care burden include occlusions of large intracranial vessels, high disease severity scores at baseline, and older age (37). Although successful recanalization of the occluded cerebral vessel during the acute ischemic event is associated with lower 3-month mortality and improved functional outcomes (78), currently approved pharmacologic treatments, including intravenous thrombolysis, have limited efficacy in these populations (910).

Neurothrombectomy devices offer many potential advantages over pharmacologic thrombolysis, including more rapid achievement of recanalization, enhanced efficacy in treating large-vessel occlusions, and a potentially lower risk for hemorrhagic events (11). Currently, 2 neurothrombectomy devices are cleared by the U.S. Food and Drug Administration (FDA) through its 510(k) process: the MERCI Retriever (Concentric Medical, Mountain View, California) and the Penumbra System (Penumbra, Alameda, California) (1215) (Figure 1). Evidence supports the benefits and safety of these devices, although studies are limited to prospective, single-group studies or retrospective designs. Various ongoing clinical trials are evaluating the effect of these devices, as well as other off-label devices, for treating acute ischemic stroke. The goal of this report is to describe the state of the evidence for use of neurothrombectomy devices for treating acute ischemic stroke.

Grahic Jump Location
Figure 1.
U.S. Food and Drug Administration–cleared devices for treating acute ischemic stroke.

Left. MERCI Retriever (Concentric Medical, Mountain View, California). Image used with permission from Concentric Medical. Right. Penumbra System (Penumbra, Alameda, California). Image used with permission from Penumbra.

Grahic Jump Location

This review is based on a technical brief produced by the University of Connecticut/Hartford Hospital Evidence-based Practice Center for the Agency for Healthcare Research and Quality (16). The full report, including literature search strategies and detailed evidence tables, is available at www.effectivehealthcare.ahrq.gov(16). We developed and followed a standard protocol for all steps of this review.

Key Questions

We formulated the following key questions with input from a technical expert panel that included experts in clinical neurology and interventional radiology, as well as a neurology association representative.

1. What neurothrombectomy devices are in clinical use or development for treating acute ischemic stroke?

2. From a systematic literature scan, provide a synthesis of trial variables, including study design, patient characteristics, comparators used, outcomes measured, and harms reported.

3. What variables may influence outcomes?

Data Sources and Searches

We searched MEDLINE, SCOPUS, the Cochrane Central Register of Controlled Trials, the Cochrane Database of Systematic Reviews, and Web of Science from the earliest possible date until May 2010. We updated the MEDLINE and Cochrane Central Register of Controlled Trials searches through November 2010. We used such terms as mechanical thrombectomy and endovascular device and text and Medical Subject Heading terms for acute ischemic stroke. The complete search strategy is available in the Appendix Table. We did not use language restrictions during the literature-identification stage. We also manually searched references from studies or reviews and searched for gray literature by using Google Scholar (Google, Menlo Park, California) and specific search terms.

Table Jump PlaceholderAppendix Table.  Search Strategies for MEDLINE and Cochrane Central Register of Controlled Trials
Study Selection

Two independent investigators screened citations at the abstract level to identify potentially relevant studies, case series, or case reports. We retrieved and examined all potentially eligible citations for eligibility. We included human studies of any design and case series or case reports that included patients with an acute ischemic stroke and reported data on at least 1 clinical effectiveness outcome or harm. Although we did not formally include duplicate reports or subgroup analyses of previously included studies in our tallies, we did review these reports for potentially pertinent information before exclusion. We extracted and qualitatively reviewed only reports published in English.

Data Extraction and Risk for Bias

Using a standardized data abstraction tool, 2 reviewers independently extracted data, with disagreement resolved by a third reviewer. The following information was obtained from each report, where applicable: author identification, year of publication, study design characteristics, study population, patient baseline characteristics, disease severity (baseline National Institutes of Health Stroke Scale [NIHSS] and baseline Thrombolysis in Myocardial Infarction [TIMI] flow grade), location of occluded artery, time from symptom onset to device deployment or angiography, use of concurrent standard medical therapies, whether outcomes assessment was blinded, and the device used. Effectiveness outcomes included recanalization as measured by TIMI flow grade after treatment (0 to 1 = no recanalization; 2 = partial recanalization; 3 = complete recanalization) or similar method, mortality, modified Rankin Scale (≤2 = good outcome; ≥3 = poor outcome), NIHSS score (including the ≥4-point decrease deemed significant by the FDA) (17), Barthel Index, and Glasgow Outcome Scale score. Harms assessed were failure to deploy the device or remove the clot (technical success); device breakage or fracture; perforation; dissection; thrombus formation proximal, adjacent, or distal to the clot site; vasospasm; or hemorrhage (including symptomatic and asymptomatic intracranial and subarachnoid hemorrhage from vessel injury and other bleeding).

We assessed the study design and classified it as a prospective, single-group study; retrospective study enrolling consecutive patients; or case series or case report. For prospective, single-group, and retrospective studies enrolling consecutive patients, we extracted data on whether outcome assessment was blinded.

Data Synthesis

The current report provides summary statistics for benefits and harms of interest. We created study density figures to summarize the information available on the effectiveness and safety of the devices. No formal quantitative synthesis (meta-analysis) was done.

Role of the Funding Source

The University of Connecticut/Hartford Hospital Evidence-based Practice Center, Hartford, Connecticut, prepared this project, with funding from the Agency for Healthcare Research and Quality. The funding source formulated the initial study questions and provided copyright release for this manuscript but did not participate in the literature search, data analysis, or interpretation of results.

Neurothrombectomy Devices in Use or Development

Five broad groupings classify neurothrombectomy devices as clot retrievers, aspiration or suction devices, snare-like devices, ultrasonography technologies, or lasers. Their advantages include avoiding or reducing the use of pharmacologic thrombolysis, allowing treatment beyond the short time frame to which intravenous recombinant tissue plasminogen activator (rt-PA) is limited, and providing more rapid recanalization than thrombolytics. Potential harms associated with navigating mechanical devices into the intracranial circulation may include direct trauma to the neurovasculature (including vasospasm, vessel dissection, perforation, or rupture) and fragmenting thrombi into previously unaffected vessels and cerebral territories (18). Although intracranial hemorrhage was an early concern about device use (19), more recent data show that its incidence is similar to that with intravenous rt-PA therapy, although differing definitions and patient populations limit these conclusions (2021). An inherent risk of the procedure includes the need for intubation and heavy sedation, which may worsen outcomes (22).

The approval process for medical devices differs from that for pharmacologic agents. The FDA Center for Device and Radiological Health handles the regulation of medical devices both premarket and postmarket (17). As such, neurothrombectomy devices are reviewed and cleared through the 510(k) premarket notification process (15, 17). For a device to receive clearance, the manufacturer must demonstrate that the new device is equal in safety and effectiveness to a Class II device that is already on the market for a particular indication. Most commonly, devices receive clearance based on nonclinical testing with little to no clinical data (17).

The MERCI Retriever and the Penumbra System are the only devices with FDA clearance for the treatment of patients with an acute ischemic stroke. The MERCI Retriever “is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke (who are ineligible for treatment with intravenous rtPA or who fail intravenous rtPA therapy)” (12, 15). The Penumbra System is used for the “revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral—M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset” (15). Other devices have FDA clearance for indications ranging from retrieval of intravascular foreign bodies to infusion of fluids into the peripheral vasculature. The In-Time (Boston Scientific, Natick, Massachusetts), Alligator (Chestnut Medical Technologies, Menlo Park, California), Amplatz Gooseneck (ev3 Endovascular, Plymouth, Minnesota), and EnSnare (Merit Medical Systems, Gainesville, Florida) devices are cleared for the retrieval of intravascular foreign objects either in general or in the peripheral vasculature and neurovasculature. The EKOS (EKOS, Bothell, Washington) and OmniWave (OmniSonics, Wilmington, Massachusetts) systems are cleared for the infusion of fluids into the peripheral vasculature. The LaTIS laser (Spectranetics, Colorado Springs, Colorado) and Oasis Thrombectomy System (Boston Scientific) are cleared for the removal of thrombi from vascular or hemodialysis access grafts. The Amplatz Thrombectomy Device (ev3 Endovascular) is cleared for dissolution of thrombi within dialysis fistulae. Data on the use of these various devices in treating patients with acute ischemic stroke are limited.

Recent and ongoing studies are evaluating the use of “retrievable” intracranial stents for managing acute ischemic stroke. They offer the ability to provide immediate recanalization, while being removable with clot trapped within the stent matrix (2328). Initial data show promise with this technology. A recent prospective, single-center pilot study reported on the safety and efficacy of a retrievable stent in 20 patients with acute stroke and large-vessel occlusion who were either refractory or ineligible for intravenous rt-PA therapy (28). The stents were deployed for 1 to 2 minutes before retrieval, with 18 of 20 patients (90%) achieving successful revascularization. Six patients (30%) had asymptomatic intracranial hemorrhage, and 2 patients (10%) had symptomatic intracranial hemorrhage (28). Ongoing trials are evaluating this line of products, which may provide a new method of restoring cerebral blood flow in patients presenting with symptoms of acute ischemic stroke who are unlikely to benefit from standard therapies.

Literature Evaluating Neurothrombectomy Devices in Acute Ischemic Stroke

We screened 1626 abstracts and evaluated 377 full-text articles (Figure 2). A total of 87 articles were included. Of these, 18 were prospective, single-group studies (2021, 2932, 3740, 4451); 7 were noncomparative retrospective studies enrolling consecutive patients (3336, 4143), and 62 were case series or case reports. Most of the studies (74%) were published in full text (2021, 2934, 3641, 4550), with the remainder (26%) available only in abstract form (3435, 4244, 51). Fifteen of 25 studies (60%) were published between 2008 and 2010 (21, 29, 3345). Only 3 of 18 (17%) prospective (38, 40, 45) and 1 of 7 (14%) retrospective (36) studies clearly stated that they used blinded outcome assessment.

The largest percentages of overall reports (40%) and prospective studies (31%) (2021, 2932) were for the MERCI Retriever. The Penumbra System was investigated in 10 reports (9%), of which 4 were prospective (3740). For off-label devices, 2 studies used the EKOS device (4546) and 1 each with Phenox (Phenox, Bochum, Germany) (44), Amplatz Gooseneck (47), AngioJet (Medrad/Possis, Warrendale, Pennsylvania) (48), EPAR (EndoVasix, Belmont, California) (49), Neuronet (Guidant, Santa Clara, California) (50), and LaTIS (51).

Studies typically enrolled patients older than 18 years with baseline NIHSS scores of 8 or greater (or ≥10) who presented within 8 hours of stroke symptom onset (or up to 24 hours for EKOS, EPAR or LaTIS if posterior circulation occlusion was identified) and had a complete or near-complete (TIMI flow grade, 0 to 1) occlusion of a treatable large intracranial vessel. Common exclusion criteria included older age, large infarction in the brain, abnormal hemostasis, severe or uncontrolled hypertension, hypoglycemia, and pregnancy. Studies also enrolled patients with contraindications to intravenous rt-PA therapy because of risk for adverse events, reporting outside a 3-hour window from symptom onset to intravenous rt-PA therapy, or whose intravenous rt-PA therapy had failed (target vessel not recanalized, determined by immediate angiography after the procedure). The exception was a study by Tomsick and colleagues (45), which evaluated the EKOS device. Along with reduced-dose intravenous rt-PA therapy, the EKOS device provided an infusion of intra-arterial thrombolytic therapy to patients within the first 3 hours of stroke symptoms.

The mean or median baseline NIHSS score ranged from 15 to 23 across studies. Mean or median age ranged from 42 to 68 years, and studies enrolled 20% to 57% female patients. Mean or median time from stroke symptom onset to either angiography or device deployment ranged from 141 to 388 minutes, which is well within the 8-hour time frame suggested by the FDA Center for Device and Radiological Health (17). The primary embolus was most frequently in an anterior vessel (14 studies enrolled >60% of patients with anterior occlusion), although some studies focused on posterior occlusions. A single study (38) included patients with occlusions in other areas, whereas 6 studies (3436, 4243, 45) were unclear about the location of the primary occlusion.

The guidance document from the FDA Center for Device and Radiological Health suggests assessing recanalization success by using the TIMI flow grade after treatment with a neurothrombectomy device (17). Each of the identified studies reported rates of recanalization (Table 1 and Figure 3). Recanalization success was typically defined as a TIMI flow grade of 2 (partial) or 3 (complete) within the target (or all) vessels on angiography after the procedure. Studies less frequently used a newer cerebral grading scale (the Thrombolysis in Cerebral Infarction scale), as well as neuroimaging measures (for example, final stroke lesion size and perfusion neuroimaging technologies). Rates of successful recanalization ranged from 43% to 78% with the MERCI Retriever (2021, 2936), 83% to 100% with the Penumbra System (3743), and 50% to 90% with the off-label devices (4451).

Table Jump PlaceholderTable 1.  Published Studies of Neurothrombectomy Devices for Acute Ischemic Stroke
Grahic Jump Location
Figure 3.
Effectiveness evidence for neurothrombectomy devices.

Darker shading represents more frequent evaluation or larger number of patients evaluated. BI = Barthel Index; C = case report or case series; GOS = Glasgow Outcome Scale; mRS = modified Rankin Scale; NIHSS = National Institutes of Health Stroke Scale; P = prospective; R = retrospective.

* Death was included if patients were followed after hospital discharge.

Grahic Jump Location

Clinical effectiveness, defined as the proportion of patients having a “good” outcome (modified Rankin Scale score, 0 to 2), was available in 17 of the 25 studies (68%). Individual rates ranged from 21% to 36% with the MERCI Retriever (2021, 3032), 20% to 48% with the Penumbra System (3743), and 15% to 60% with the off-label devices (4650). Seventeen of 25 studies (68%) reported mortality rates ranging from 29% to 44% with the MERCI Retriever (2021, 3032), 11% to 45% with the Penumbra System (3743), and 0% to 38% with the off-label devices (4650).

Incidences of intracranial hemorrhage, both symptomatic and asymptomatic, and vessel perforation or dissection were the most commonly reported potential harms (Figure 4). Symptomatic intracranial hemorrhage was reported in 16 of 25 studies (64%). Rates ranged from 0% to 10% with the MERCI Retriever (2021, 3032), 0% to 11% with the Penumbra System (3843), and 6% to 25% with the off-label devices (4549). Thirteen studies (52%) reported the incidence of asymptomatic intracranial hemorrhage. Rates ranged from 28% to 43% with the MERCI Retriever (2021, 3032), 1% to 30% with the Penumbra System (3843), and 6% to 17% with the off-label devices (4749). Incidence of target vessel perforation or dissection was documented less often, with only 11 of 25 studies (44%) reporting data. Rates ranged from 0% to 7% with the MERCI Retriever (20, 3031) and 0% to 5% with the Penumbra System (38, 4041), and no events occurred in the 5 off-label studies that reported the outcome (4549).

Grahic Jump Location
Figure 4.
Safety end point evidence for neurothrombectomy devices.

Darker shading represents more frequent evaluation or larger number of patients evaluated. AICH = asymptomatic intracranial hemorrhage; C = case report or case series; P = prospective; R = retrospective; SICH = symptomatic intracranial hemorrhage.

Grahic Jump Location
Variables That May Affect Clinical Outcomes

We determined the association between predictor variables and selected outcomes (effectiveness and harms) by using multivariate data from prospective and retrospective studies (7, 2021, 2627, 52). We classified predictors by whether they were suggested to have beneficial (and statistically significant), harmful (and statistically significant), or indeterminate (not statistically significant regardless of effect direction) associations with outcomes. These include various demographic, comorbid disease, stroke severity, and stroke treatment variables. Studies reporting these predictors were limited to those investigating the MERCI Retriever and Penumbra System.

Achievement of successful recanalization was the only variable that was predictive of good outcomes, although concomitant use of intra-arterial thrombolytics was predictive of achieving recanalization (20, 26, 52). Variables that predicted worse outcomes included older age (20, 52), higher baseline systolic blood pressure (20, 52), higher baseline NIHSS scores (20, 26, 52), history of stroke (26), longer procedure duration (20, 26), and documented occlusions of the internal cerebral artery (20, 26, 52). Variables linked to outcomes in patients with acute ischemic stroke included the presence of collateral circulation, lesion volume, and cerebral perfusion pressure (53). Of note, recanalization was the only variable found to be predictive of clinical benefit as well as lower mortality.

We did not identify studies assessing the relationship between the training of interventionalists and outcomes in patients treated with neurothrombectomy devices. However, studies of other emerging stroke technologies from the past 20 years have suggested that inadequate physician training and experience can adversely affect clinical outcomes (54). Two reports detail the minimum training requirements for persons performing neuroendovascular procedures (including neurothrombectomy devices) in patients with acute ischemic stroke and the performance standards that should be adopted to assess outcomes (5455).

The natural history and poor clinical outcomes seen in patients with acute ischemic strokes have been well documented (3, 7, 47). The advent of neurothrombectomy devices to provide adequate recanalization has created a new option for managing these patients. These technologies are particularly intriguing given the data showing significantly improved outcomes with successful and early recanalization of an occluded cerebral vessel (7, 56). Currently, only 2 neurothrombectomy devices, the MERCI Retriever and Penumbra System, are cleared by the FDA to treat patients with acute ischemic stroke. The comparative effectiveness of these devices remains unstudied.

Our report is the most up-to-date review of the literature on neurothrombectomy devices. A prior meta-analysis, which included studies published up to 2006, found that more patients who received neurothrombectomy devices (34.5%) had good functional recovery (modified Rankin Scale score ≤2) compared with a patient cohort matched for age, sex, and NIHSS score (10.7%) (19). Although intriguing, this analysis is fraught with limitations, including significant heterogeneity within the neurothrombectomy cohort, comparison with a historical control (albeit matched for important variables), and lack of participant randomization.

A continuing area of uncertainty is the patient population that is most likely to benefit from use of neurothrombectomy devices. Indeed, many published studies evaluating FDA-cleared devices have enrolled patients who presented within 8 hours of onset of ischemic stroke symptoms, had high baseline NIHSS scores, and were ineligible to receive intravenous rt-PA therapy or in whom intravenous rt-PA therapy had failed (3, 7, 47, 5758). In addition, many of the patients had occlusions of the large intracranial vessels, which respond poorly to intravenous rt-PA therapy (10, 59). Future randomized, controlled trials should incorporate advanced imaging techniques to better identify patients who are most likely to benefit from neurothrombectomy devices, as well as to evaluate the performance of these devices at varying levels of occlusion. Ongoing clinical trials, including the MR RESCUE (Mechanical Retrieval and Recanalization of Stroke Clots Using Embolectomy) study, are evaluating the utility of various imaging techniques to better identify which patients are most likely to benefit from neurothrombectomy device use. These emerging data should help guide clinicians in choosing the most appropriate treatment of acute ischemic strokes.

Because the literature base of evaluations of neurothrombectomy devices in acute ischemic stroke is still in its infancy, identifying gaps in current knowledge and guiding future research efforts are paramount. Currently, 11 ongoing studies evaluating at least 1 neurothrombectomy device in acute ischemic stroke are listed in ClinicalTrials.gov or are mentioned in previous review articles (53) (Table 2). Seven of these studies have randomized, controlled designs, with projected enrollment ranging from 20 to 900 participants. The other 4 studies have prospective, observational designs, with enrollment ranging from 200 to 2000 projected participants. Of note, the SWIFT (SOLITAIR FR With the Intention for Thrombectomy) Study will be the first to evaluate the comparative effectiveness of 2 different technologies, the SOLITAIRE FR (ev3 Endovascular) and the MERCI Retriever, by using a prospective, randomized method in patients with acute ischemic stroke.

Table Jump PlaceholderTable 2.  Summary of Ongoing Studies of Neurothrombectomy Devices for Ischemic Stroke

Our literature search did not identify any reports evaluating the effect of neurothrombectomy devices on health-related quality-of-life outcomes. We identified only 1 economic evaluation of neurothrombectomy devices: a Markov model that evaluated the cost and effectiveness of mechanical thrombectomy compared with standard medical therapy in patients who were ineligible to receive intravenous rt-PA therapy (60). The ongoing IMS (Interventional Management of Stroke) III Trial, which is designed to evaluate combination use of intravenous and intra-arterial strategies (including use of the MERCI Retriever or EKOS device) versus intravenous rt-PA therapy alone, plans to measure both health-related quality-of-life and pharmacoeconomic outcomes (6162).

Currently available neurothrombectomy devices offer intriguing treatment options in patients with acute ischemic stroke, although a paucity of high-quality research currently exists. Further research is needed on the topic, including randomized, controlled trials to determine the optimal device to use and the patient populations that are most likely to benefit from its use. The results of ongoing studies, some of which are using randomized, controlled designs, are eagerly anticipated. In addition, future studies should investigate whether neurothrombectomy devices affect final health outcomes associated with stroke rather than improving recanalization alone when compared with contemporary controls.

Lopez AD, Mathers CD, Ezzati M, Jamison DT, Murray CJ.  Global and regional burden of disease and risk factors, 2001: systematic analysis of population health data. Lancet. 2006; 367:1747-57.
PubMed
CrossRef
 
Lloyd-Jones D, Adams R, Carnethon M, DeSimone G, Ferguson TB, Flegal K, et al. American Heart Association Statistics Committee and Stroke Statistics Subcommittee.  Heart disease and stroke statistics—2009 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Circulation. 2009; 119:480-6.
PubMed
 
Smith WS, Tsao JW, Billings ME, Johnston SC, Hemphill JC 3rd, Bonovich DC. et al.  Prognostic significance of angiographically confirmed large vessel intracranial occlusion in patients presenting with acute brain ischemia. Neurocrit Care. 2006; 4:14-7.
PubMed
 
Centers for Disease Control and Prevention (CDC).  Prevalence of disabilities and associated health conditions among adults—United States, 1999. MMWR Morb Mortal Wkly Rep. 2001; 50:120-5.
PubMed
 
Centers for Disease Control and Prevention (CDC).  Outpatient rehabilitation among stroke survivors—21 States and the District of Columbia, 2005. MMWR Morb Mortal Wkly Rep. 2007; 56:504-7.
PubMed
 
Fischer U, Arnold M, Nedeltchev K, Brekenfeld C, Ballinari P, Remonda L. et al.  NIHSS score and arteriographic findings in acute ischemic stroke. Stroke. 2005; 36:2121-5.
PubMed
 
Nogueira RG, Liebeskind DS, Sung G, Duckwiler G, Smith WS, MERCI.  Predictors of good clinical outcomes, mortality, and successful revascularization in patients with acute ischemic stroke undergoing thrombectomy: pooled analysis of the Mechanical Embolus Removal in Cerebral Ischemia (MERCI) and Multi MERCI Trials. Stroke. 2009; 40:3777-83.
PubMed
 
Rha JH, Saver JL.  The impact of recanalization on ischemic stroke outcome: a meta-analysis. Stroke. 2007; 38:967-73.
PubMed
 
Zangerle A, Kiechl S, Spiegel M, Furtner M, Knoflach M, Werner P. et al.  Recanalization after thrombolysis in stroke patients: predictors and prognostic implications. Neurology. 2007; 68:39-44.
PubMed
 
Wolpert SM, Bruckmann H, Greenlee R, Wechsler L, Pessin MS, del Zoppo GJ.  Neuroradiologic evaluation of patients with acute stroke treated with recombinant tissue plasminogen activator. The rt-PA Acute Stroke Study Group. AJNR Am J Neuroradiol. 1993; 14:3-13.
PubMed
 
Thomassen L, Bakke SJ.  Endovascular reperfusion therapy in acute ischaemic stroke. Acta Neurol Scand Suppl. 2007; 187:22-9.
PubMed
 
Becker KJ, Brott TG.  Approval of the MERCI clot retriever: a critical view. Stroke. 2005; 36:400-3.
PubMed
 
Nogueira RG, Schwamm LH, Hirsch JA.  Endovascular approaches to acute stroke, part 1: Drugs, devices, and data. AJNR Am J Neuroradiol. 2009; 30:649-61.
PubMed
 
Lutsep HL.  Mechanical endovascular recanalization therapies. Curr Opin Neurol. 2008; 21:70-5.
PubMed
 
U.S. Food and Drug Administration 510(k) Premarket Notification. Accessed atwww.accessdata.fda.gov/scrIpts/cdrh/cfdocs/cfPMN/pmn.cfmon 25 August 2010.
 
Baker WL, Colby JA, Tongbram V, Talati RA, Silverman IE, White CM, et al.  Neurothrombectomy devices for treatment of acute ischemic stroke. (Prepared by the University of Connecticut/Hartford Hospital Evidence-based Practice Center under contract 290-2007-10067-I with the Agency for Healthcare Research and Quality.) Rockville, MD: Agency for Healthcare Research and Quality; 2010. Accessed atwww.effectivehealthcare.ahrq.govon 16 December 2010.
 
Guidance for Industry and FDA Staff: Pre-Clinical and Clinical Studies for Neurothrombectomy Devices. Silver Spring, MD: U.S. Department of Health and Human Services, Food and Drug Administration, Center for Devices and Radiological Health; 18 June 2007. Accessed at:www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm071403.htmon 25 August 2010.
 
Gupta R.  Arterial vasospasm during mechanical thrombectomy for acute stroke. J Neuroimaging. 2009; 19:61-4.
PubMed
 
Stead LG, Gilmore RM, Bellolio MF, Rabinstein AA, Decker WW.  Percutaneous clot removal devices in acute ischemic stroke: a systematic review and meta-analysis. Arch Neurol. 2008; 65:1024-30.
PubMed
 
Smith WS, Sung G, Starkman S, Saver JL, Kidwell CS, Gobin YP, et al. MERCI Trial Investigators.  Safety and efficacy of mechanical embolectomy in acute ischemic stroke: results of the MERCI trial. Stroke. 2005; 36:1432-8.
PubMed
 
Smith WS, Sung G, Saver J, Budzik R, Duckwiler G, Liebeskind DS, et al. Multi MERCI Investigators.  Mechanical thrombectomy for acute ischemic stroke: final results of the Multi MERCI trial. Stroke. 2008; 39:1205-12.
PubMed
 
Nichols C, Carrozzella J, Yeatts S, Tomsick T, Broderick J, Khatri P.  Is peri-procedural sedation during acute stroke therapy associated with poorer functional outcomes? J Neurointerv Surg. 2010; 2:67-70.
PubMed
 
Kelly ME, Furlan AJ, Fiorella D.  Recanalization of an acute middle cerebral artery occlusion using a self-expanding, reconstrainable, intracranial microstent as a temporary endovascular bypass. Stroke. 2008; 39:1770-3.
PubMed
 
Suh SH, Lee KY, Hong CK, Kim BM, Kim CH, Chung TS. et al.  Temporary stenting and retrieval of the self-expandable, intracranial stent in acute middle cerebral artery occlusion. Neuroradiology. 2009; 51:541-4.
PubMed
 
Hauck EF, Mocco J, Snyder KV, Levy EI.  Temporary endovascular bypass: a novel treatment for acute stroke. AJNR Am J Neuroradiol. 2009; 30:1532-3.
PubMed
 
Castaño C, Serena J, Dávalos A.  Use of the new Solitaire (TM) AB device for mechanical thrombectomy when Merci Clot Retriever has failed to remove the clot. A case report. Interv Neuroradiol. 2009; 15:209-14.
PubMed
 
Henkes H, Liebig T, Miloslavski E, Guethe T, Schmid E, Wiethoelter H.  Endovascular acute ischemic stroke treatment using the self-expanding and fully retrievable Solitaire stent [Abstract]. Stroke. 2009; 40:247-8.
 
Castaño C, Dorado L, Guerrero C, Millán M, Gomis M, Perez de la Ossa N. et al.  Mechanical thrombectomy with the Solitaire AB device in large artery occlusions of the anterior circulation: a pilot study [Letter]. Stroke. 2010; 41:1836-40.
PubMed
 
Loh Y, Jahan R, McArthur DL, Shi ZS, Gonzalez NR, Duckwiler GR. et al.  Recanalization rates decrease with increasing thrombectomy attempts. AJNR Am J Neuroradiol. 2010; 31:935-9.
PubMed
 
Devlin TG, Baxter BW, Feintuch TA, Desbiens .  The Merci Retrieval System for acute stroke: the Southeast Regional Stroke Center experience. Neurocrit Care. 2007; 6:11-21.
PubMed
 
Kim D, Jahan R, Starkman S, Abolian A, Kidwell CS, Vinuela F. et al.  Endovascular mechanical clot retrieval in a broad ischemic stroke cohort. AJNR Am J Neuroradiol. 2006; 27:2048-52.
PubMed
 
Gobin YP, Starkman S, Duckwiler GR, Grobelny T, Kidwell CS, Jahan R. et al.  MERCI 1: a phase 1 study of Mechanical Embolus Removal in Cerebral Ischemia. Stroke. 2004; 35:2848-54.
PubMed
 
Lin R, Vora N, Zaidi S, Aleu A, Jankowitz B, Thomas A. et al.  Mechanical approaches combined with intra-arterial pharmacological therapy are associated with higher recanalization rates than either intervention alone in revascularization of acute carotid terminus occlusion. Stroke. 2009; 40:2092-7.
PubMed
 
Jo KD, Saver JL, Starkman S, Kim D, Ali LK, Ovbiagele B, et al.  Predictors of recanalization with mechanical thrombectomy for acute ischemic stroke [Abstract]. Stroke. 2008;39:599.
 
Madison MT, Goddard JK. III, Lassig JP, Meyers ME . Single center experience with combined (thrombolytics and mechanical thrombectomy) intra-arterial intervention in 112 consecutive patients with angiographically confirmed large vessel thromboembolic ischemic stroke (1-6 hours post symptom onset) from 2003 to 2007 [Abstract]. Stroke. 2008;39:547.
 
Kidwell CS, Latour L, Saver JL, Alger JR, Starkman S, Duckwiler G, et al. UCLA Thrombolysis Investigators.  Thrombolytic toxicity: blood brain barrier disruption in human ischemic stroke. Cerebrovasc Dis. 2008; 25:338-43.
PubMed
 
Kulcsár Z, Bonvin C, Pereira VM, Altrichter S, Yilmaz H, Lövblad KO. et al.  Penumbra system: a novel mechanical thrombectomy device for large-vessel occlusions in acute stroke. AJNR Am J Neuroradiol. 2010; 31:628-33.
PubMed
 
Penumbra Pivotal Stroke Trial Investigators.  The Penumbra Pivotal Stroke Trial: safety and effectiveness of a new generation of mechanical devices for clot removal in intracranial large vessel occlusive disease. Stroke. 2009; 40:2761-8.
PubMed
 
Grunwald IQ, Walter S, Papanagiotou P, Krick C, Hartmann K, Dautermann A. et al.  Revascularization in acute ischaemic stroke using the penumbra system: the first single center experience. Eur J Neurol. 2009; 16:1210-6.
PubMed
 
Bose A, Henkes H, Alfke K, Reith W, Mayer TE, Berlis A, et al. Penumbra Phase 1 Stroke Trial Investigators.  The Penumbra System: a mechanical device for the treatment of acute stroke due to thromboembolism. AJNR Am J Neuroradiol. 2008; 29:1409-13.
PubMed
 
Struffert T, Köhrmann M, Engelhorn T, Nowe T, Richter G, Schellinger PD. et al.  Penumbra Stroke System as an “add-on” for the treatment of large vessel occlusive disease following thrombolysis: first results. Eur Radiol. 2009; 19:2286-93.
PubMed
 
Tarr R, Alfke K, Stingele R, Jansen O, Frei D, Bellon R, et al.  Initial post-market experience of the Penumbra System: revascularization of large vessel occlusion in acute ischemic stroke in the United States and Europe [Abstract]. Stroke. 2009;40:e151.
 
Frei D, Bellon R.  Mechanical thrombectomy in acute stroke patients who were refractory to intravenous thrombolytic therapy [Abstract]. Stroke. 2009;40:E250.
 
Liebig T, Reinartz J, Guethe T, Roth C, Milosiavski M.  Early clinical experiences with a new thrombectomy device for the treatment of ischemic stroke [Abstract]. Stroke. 2008; 39:608-9.
 
Tomsick T, Broderick J, Carrozella J, Khatri P, Hill M, Palesch Y, et al. Interventional Management of Stroke II Investigators.  Revascularization results in the Interventional Management of Stroke II trial. AJNR Am J Neuroradiol. 2008; 29:582-7.
PubMed
 
Mahon BR, Nesbit GM, Barnwell SL, Clark W, Marotta TR, Weill A. et al.  North American clinical experience with the EKOS MicroLysUS infusion catheter for the treatment of embolic stroke. AJNR Am J Neuroradiol. 2003; 24:534-8.
PubMed
 
González A, Mayol A, Martínez E, González-Marcos JR, Gil-Peralta A.  Mechanical thrombectomy with snare in patients with acute ischemic stroke. Neuroradiology. 2007; 49:365-72.
PubMed
 
Mayer TE, Hamann GF, Schulte-Altedorneburg G, Brückmann H.  Treatment of vertebrobasilar occlusion by using a coronary waterjet thrombectomy device: a pilot study. AJNR Am J Neuroradiol. 2005; 26:1389-94.
PubMed
 
Berlis A, Lutsep H, Barnwell S, Norbash A, Wechsler L, Jungreis CA. et al.  Mechanical thrombolysis in acute ischemic stroke with endovascular photoacoustic recanalization. Stroke. 2004; 35:1112-6.
PubMed
 
Mayer TE, Hamann GF, Brueckmann HJ.  Treatment of basilar artery embolism with a mechanical extraction device: necessity of flow reversal. Stroke. 2002; 33:2232-5.
PubMed
 
Clark RA, Bryant AL, Pua Y, McCrory P, Bennell K, Hunt M.  Validity and reliability of the Nintendo Wii Balance Board for assessment of standing balance. Gait Posture. 2010; 31:307-10.
PubMed
 
Nogueira RG, Smith WS, MERCI and Multi MERCI Writing Committee.  Safety and efficacy of endovascular thrombectomy in patients with abnormal hemostasis: pooled analysis of the MERCI and multi MERCI trials. Stroke. 2009; 40:516-22.
PubMed
 
Nogueira RG, Yoo AJ, Buonanno FS, Hirsch JA.  Endovascular approaches to acute stroke, part 2: a comprehensive review of studies and trials. AJNR Am J Neuroradiol. 2009; 30:859-75.
PubMed
 
Meyers PM, Schumacher HC, Alexander MJ, Derdeyn CP, Furlan AJ, Higashida RT, et al. American Academy of Neurology.  Performance and training standards for endovascular ischemic stroke treatment. J Neurosurg. 2010; 113:149-52.
PubMed
 
Connors JJ 3rd, Sacks D, Black CM, McIff EB, Stallmeyer MJ, Cole JW, et al. Society of Interventional Radiology.  Training guidelines for intra-arterial catheter-directed treatment of acute ischemic stroke: a statement from a special writing group of the Society of Interventional Radiology. J Vasc Interv Radiol. 2009; 20:1507-22.
PubMed
 
Khatri P, Abruzzo T, Yeatts SD, Nichols C, Broderick JP, Tomsick TA, IMS I and II Investigators.  Good clinical outcome after ischemic stroke with successful revascularization is time-dependent. Neurology. 2009; 73:1066-72.
PubMed
 
Smith WS, Lev MH, English JD, Camargo EC, Chou M, Johnston SC. et al.  Significance of large vessel intracranial occlusion causing acute ischemic stroke and TIA. Stroke. 2009; 40:3834-40.
PubMed
 
Josephson SA, Saver JL, Smith WS, Merci and Multi Merci Investigators.  Comparison of mechanical embolectomy and intraarterial thrombolysis in acute ischemic stroke within the MCA: MERCI and Multi MERCI compared to PROACT II. Neurocrit Care. 2009; 10:43-9.
PubMed
 
Alexandrov AV, Grotta JC.  Arterial reocclusion in stroke patients treated with intravenous tissue plasminogen activator. Neurology. 2002; 59:862-7.
PubMed
 
Gralla J, Schroth G, Remonda L, Nedeltchev K, Slotboom J, Brekenfeld C.  Mechanical thrombectomy for acute ischemic stroke: thrombus-device interaction, efficiency, and complications in vivo. Stroke. 2006; 37:3019-24.
PubMed
 
Khatri P, Hill MD, Palesch YY, Spilker J, Jauch EC, Carrozzella JA, et al. Interventional Management of Stroke III Investigators.  Methodology of the Interventional Management of Stroke III Trial. Int J Stroke. 2008; 3:130-7.
PubMed
 
Mauldin PD, Simpson KN, Palesch YY, Spilker JS, Hill MD, Khatri P, et al. Interventional Management of Stroke III Investigators.  Design of the economic evaluation for the Interventional Management of Stroke (III) trial. Int J Stroke. 2008; 3:138-44.
PubMed
 

Figures

Grahic Jump Location
Figure 1.
U.S. Food and Drug Administration–cleared devices for treating acute ischemic stroke.

Left. MERCI Retriever (Concentric Medical, Mountain View, California). Image used with permission from Concentric Medical. Right. Penumbra System (Penumbra, Alameda, California). Image used with permission from Penumbra.

Grahic Jump Location
Grahic Jump Location
Figure 3.
Effectiveness evidence for neurothrombectomy devices.

Darker shading represents more frequent evaluation or larger number of patients evaluated. BI = Barthel Index; C = case report or case series; GOS = Glasgow Outcome Scale; mRS = modified Rankin Scale; NIHSS = National Institutes of Health Stroke Scale; P = prospective; R = retrospective.

* Death was included if patients were followed after hospital discharge.

Grahic Jump Location
Grahic Jump Location
Figure 4.
Safety end point evidence for neurothrombectomy devices.

Darker shading represents more frequent evaluation or larger number of patients evaluated. AICH = asymptomatic intracranial hemorrhage; C = case report or case series; P = prospective; R = retrospective; SICH = symptomatic intracranial hemorrhage.

Grahic Jump Location

Tables

Table Jump PlaceholderAppendix Table.  Search Strategies for MEDLINE and Cochrane Central Register of Controlled Trials
Table Jump PlaceholderTable 1.  Published Studies of Neurothrombectomy Devices for Acute Ischemic Stroke
Table Jump PlaceholderTable 2.  Summary of Ongoing Studies of Neurothrombectomy Devices for Ischemic Stroke

References

Lopez AD, Mathers CD, Ezzati M, Jamison DT, Murray CJ.  Global and regional burden of disease and risk factors, 2001: systematic analysis of population health data. Lancet. 2006; 367:1747-57.
PubMed
CrossRef
 
Lloyd-Jones D, Adams R, Carnethon M, DeSimone G, Ferguson TB, Flegal K, et al. American Heart Association Statistics Committee and Stroke Statistics Subcommittee.  Heart disease and stroke statistics—2009 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Circulation. 2009; 119:480-6.
PubMed
 
Smith WS, Tsao JW, Billings ME, Johnston SC, Hemphill JC 3rd, Bonovich DC. et al.  Prognostic significance of angiographically confirmed large vessel intracranial occlusion in patients presenting with acute brain ischemia. Neurocrit Care. 2006; 4:14-7.
PubMed
 
Centers for Disease Control and Prevention (CDC).  Prevalence of disabilities and associated health conditions among adults—United States, 1999. MMWR Morb Mortal Wkly Rep. 2001; 50:120-5.
PubMed
 
Centers for Disease Control and Prevention (CDC).  Outpatient rehabilitation among stroke survivors—21 States and the District of Columbia, 2005. MMWR Morb Mortal Wkly Rep. 2007; 56:504-7.
PubMed
 
Fischer U, Arnold M, Nedeltchev K, Brekenfeld C, Ballinari P, Remonda L. et al.  NIHSS score and arteriographic findings in acute ischemic stroke. Stroke. 2005; 36:2121-5.
PubMed
 
Nogueira RG, Liebeskind DS, Sung G, Duckwiler G, Smith WS, MERCI.  Predictors of good clinical outcomes, mortality, and successful revascularization in patients with acute ischemic stroke undergoing thrombectomy: pooled analysis of the Mechanical Embolus Removal in Cerebral Ischemia (MERCI) and Multi MERCI Trials. Stroke. 2009; 40:3777-83.
PubMed
 
Rha JH, Saver JL.  The impact of recanalization on ischemic stroke outcome: a meta-analysis. Stroke. 2007; 38:967-73.
PubMed
 
Zangerle A, Kiechl S, Spiegel M, Furtner M, Knoflach M, Werner P. et al.  Recanalization after thrombolysis in stroke patients: predictors and prognostic implications. Neurology. 2007; 68:39-44.
PubMed
 
Wolpert SM, Bruckmann H, Greenlee R, Wechsler L, Pessin MS, del Zoppo GJ.  Neuroradiologic evaluation of patients with acute stroke treated with recombinant tissue plasminogen activator. The rt-PA Acute Stroke Study Group. AJNR Am J Neuroradiol. 1993; 14:3-13.
PubMed
 
Thomassen L, Bakke SJ.  Endovascular reperfusion therapy in acute ischaemic stroke. Acta Neurol Scand Suppl. 2007; 187:22-9.
PubMed
 
Becker KJ, Brott TG.  Approval of the MERCI clot retriever: a critical view. Stroke. 2005; 36:400-3.
PubMed
 
Nogueira RG, Schwamm LH, Hirsch JA.  Endovascular approaches to acute stroke, part 1: Drugs, devices, and data. AJNR Am J Neuroradiol. 2009; 30:649-61.
PubMed
 
Lutsep HL.  Mechanical endovascular recanalization therapies. Curr Opin Neurol. 2008; 21:70-5.
PubMed
 
U.S. Food and Drug Administration 510(k) Premarket Notification. Accessed atwww.accessdata.fda.gov/scrIpts/cdrh/cfdocs/cfPMN/pmn.cfmon 25 August 2010.
 
Baker WL, Colby JA, Tongbram V, Talati RA, Silverman IE, White CM, et al.  Neurothrombectomy devices for treatment of acute ischemic stroke. (Prepared by the University of Connecticut/Hartford Hospital Evidence-based Practice Center under contract 290-2007-10067-I with the Agency for Healthcare Research and Quality.) Rockville, MD: Agency for Healthcare Research and Quality; 2010. Accessed atwww.effectivehealthcare.ahrq.govon 16 December 2010.
 
Guidance for Industry and FDA Staff: Pre-Clinical and Clinical Studies for Neurothrombectomy Devices. Silver Spring, MD: U.S. Department of Health and Human Services, Food and Drug Administration, Center for Devices and Radiological Health; 18 June 2007. Accessed at:www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm071403.htmon 25 August 2010.
 
Gupta R.  Arterial vasospasm during mechanical thrombectomy for acute stroke. J Neuroimaging. 2009; 19:61-4.
PubMed
 
Stead LG, Gilmore RM, Bellolio MF, Rabinstein AA, Decker WW.  Percutaneous clot removal devices in acute ischemic stroke: a systematic review and meta-analysis. Arch Neurol. 2008; 65:1024-30.
PubMed
 
Smith WS, Sung G, Starkman S, Saver JL, Kidwell CS, Gobin YP, et al. MERCI Trial Investigators.  Safety and efficacy of mechanical embolectomy in acute ischemic stroke: results of the MERCI trial. Stroke. 2005; 36:1432-8.
PubMed
 
Smith WS, Sung G, Saver J, Budzik R, Duckwiler G, Liebeskind DS, et al. Multi MERCI Investigators.  Mechanical thrombectomy for acute ischemic stroke: final results of the Multi MERCI trial. Stroke. 2008; 39:1205-12.
PubMed
 
Nichols C, Carrozzella J, Yeatts S, Tomsick T, Broderick J, Khatri P.  Is peri-procedural sedation during acute stroke therapy associated with poorer functional outcomes? J Neurointerv Surg. 2010; 2:67-70.
PubMed
 
Kelly ME, Furlan AJ, Fiorella D.  Recanalization of an acute middle cerebral artery occlusion using a self-expanding, reconstrainable, intracranial microstent as a temporary endovascular bypass. Stroke. 2008; 39:1770-3.
PubMed
 
Suh SH, Lee KY, Hong CK, Kim BM, Kim CH, Chung TS. et al.  Temporary stenting and retrieval of the self-expandable, intracranial stent in acute middle cerebral artery occlusion. Neuroradiology. 2009; 51:541-4.
PubMed
 
Hauck EF, Mocco J, Snyder KV, Levy EI.  Temporary endovascular bypass: a novel treatment for acute stroke. AJNR Am J Neuroradiol. 2009; 30:1532-3.
PubMed
 
Castaño C, Serena J, Dávalos A.  Use of the new Solitaire (TM) AB device for mechanical thrombectomy when Merci Clot Retriever has failed to remove the clot. A case report. Interv Neuroradiol. 2009; 15:209-14.
PubMed
 
Henkes H, Liebig T, Miloslavski E, Guethe T, Schmid E, Wiethoelter H.  Endovascular acute ischemic stroke treatment using the self-expanding and fully retrievable Solitaire stent [Abstract]. Stroke. 2009; 40:247-8.
 
Castaño C, Dorado L, Guerrero C, Millán M, Gomis M, Perez de la Ossa N. et al.  Mechanical thrombectomy with the Solitaire AB device in large artery occlusions of the anterior circulation: a pilot study [Letter]. Stroke. 2010; 41:1836-40.
PubMed
 
Loh Y, Jahan R, McArthur DL, Shi ZS, Gonzalez NR, Duckwiler GR. et al.  Recanalization rates decrease with increasing thrombectomy attempts. AJNR Am J Neuroradiol. 2010; 31:935-9.
PubMed
 
Devlin TG, Baxter BW, Feintuch TA, Desbiens .  The Merci Retrieval System for acute stroke: the Southeast Regional Stroke Center experience. Neurocrit Care. 2007; 6:11-21.
PubMed
 
Kim D, Jahan R, Starkman S, Abolian A, Kidwell CS, Vinuela F. et al.  Endovascular mechanical clot retrieval in a broad ischemic stroke cohort. AJNR Am J Neuroradiol. 2006; 27:2048-52.
PubMed
 
Gobin YP, Starkman S, Duckwiler GR, Grobelny T, Kidwell CS, Jahan R. et al.  MERCI 1: a phase 1 study of Mechanical Embolus Removal in Cerebral Ischemia. Stroke. 2004; 35:2848-54.
PubMed
 
Lin R, Vora N, Zaidi S, Aleu A, Jankowitz B, Thomas A. et al.  Mechanical approaches combined with intra-arterial pharmacological therapy are associated with higher recanalization rates than either intervention alone in revascularization of acute carotid terminus occlusion. Stroke. 2009; 40:2092-7.
PubMed
 
Jo KD, Saver JL, Starkman S, Kim D, Ali LK, Ovbiagele B, et al.  Predictors of recanalization with mechanical thrombectomy for acute ischemic stroke [Abstract]. Stroke. 2008;39:599.
 
Madison MT, Goddard JK. III, Lassig JP, Meyers ME . Single center experience with combined (thrombolytics and mechanical thrombectomy) intra-arterial intervention in 112 consecutive patients with angiographically confirmed large vessel thromboembolic ischemic stroke (1-6 hours post symptom onset) from 2003 to 2007 [Abstract]. Stroke. 2008;39:547.
 
Kidwell CS, Latour L, Saver JL, Alger JR, Starkman S, Duckwiler G, et al. UCLA Thrombolysis Investigators.  Thrombolytic toxicity: blood brain barrier disruption in human ischemic stroke. Cerebrovasc Dis. 2008; 25:338-43.
PubMed
 
Kulcsár Z, Bonvin C, Pereira VM, Altrichter S, Yilmaz H, Lövblad KO. et al.  Penumbra system: a novel mechanical thrombectomy device for large-vessel occlusions in acute stroke. AJNR Am J Neuroradiol. 2010; 31:628-33.
PubMed
 
Penumbra Pivotal Stroke Trial Investigators.  The Penumbra Pivotal Stroke Trial: safety and effectiveness of a new generation of mechanical devices for clot removal in intracranial large vessel occlusive disease. Stroke. 2009; 40:2761-8.
PubMed
 
Grunwald IQ, Walter S, Papanagiotou P, Krick C, Hartmann K, Dautermann A. et al.  Revascularization in acute ischaemic stroke using the penumbra system: the first single center experience. Eur J Neurol. 2009; 16:1210-6.
PubMed
 
Bose A, Henkes H, Alfke K, Reith W, Mayer TE, Berlis A, et al. Penumbra Phase 1 Stroke Trial Investigators.  The Penumbra System: a mechanical device for the treatment of acute stroke due to thromboembolism. AJNR Am J Neuroradiol. 2008; 29:1409-13.
PubMed
 
Struffert T, Köhrmann M, Engelhorn T, Nowe T, Richter G, Schellinger PD. et al.  Penumbra Stroke System as an “add-on” for the treatment of large vessel occlusive disease following thrombolysis: first results. Eur Radiol. 2009; 19:2286-93.
PubMed
 
Tarr R, Alfke K, Stingele R, Jansen O, Frei D, Bellon R, et al.  Initial post-market experience of the Penumbra System: revascularization of large vessel occlusion in acute ischemic stroke in the United States and Europe [Abstract]. Stroke. 2009;40:e151.
 
Frei D, Bellon R.  Mechanical thrombectomy in acute stroke patients who were refractory to intravenous thrombolytic therapy [Abstract]. Stroke. 2009;40:E250.
 
Liebig T, Reinartz J, Guethe T, Roth C, Milosiavski M.  Early clinical experiences with a new thrombectomy device for the treatment of ischemic stroke [Abstract]. Stroke. 2008; 39:608-9.
 
Tomsick T, Broderick J, Carrozella J, Khatri P, Hill M, Palesch Y, et al. Interventional Management of Stroke II Investigators.  Revascularization results in the Interventional Management of Stroke II trial. AJNR Am J Neuroradiol. 2008; 29:582-7.
PubMed
 
Mahon BR, Nesbit GM, Barnwell SL, Clark W, Marotta TR, Weill A. et al.  North American clinical experience with the EKOS MicroLysUS infusion catheter for the treatment of embolic stroke. AJNR Am J Neuroradiol. 2003; 24:534-8.
PubMed
 
González A, Mayol A, Martínez E, González-Marcos JR, Gil-Peralta A.  Mechanical thrombectomy with snare in patients with acute ischemic stroke. Neuroradiology. 2007; 49:365-72.
PubMed
 
Mayer TE, Hamann GF, Schulte-Altedorneburg G, Brückmann H.  Treatment of vertebrobasilar occlusion by using a coronary waterjet thrombectomy device: a pilot study. AJNR Am J Neuroradiol. 2005; 26:1389-94.
PubMed
 
Berlis A, Lutsep H, Barnwell S, Norbash A, Wechsler L, Jungreis CA. et al.  Mechanical thrombolysis in acute ischemic stroke with endovascular photoacoustic recanalization. Stroke. 2004; 35:1112-6.
PubMed
 
Mayer TE, Hamann GF, Brueckmann HJ.  Treatment of basilar artery embolism with a mechanical extraction device: necessity of flow reversal. Stroke. 2002; 33:2232-5.
PubMed
 
Clark RA, Bryant AL, Pua Y, McCrory P, Bennell K, Hunt M.  Validity and reliability of the Nintendo Wii Balance Board for assessment of standing balance. Gait Posture. 2010; 31:307-10.
PubMed
 
Nogueira RG, Smith WS, MERCI and Multi MERCI Writing Committee.  Safety and efficacy of endovascular thrombectomy in patients with abnormal hemostasis: pooled analysis of the MERCI and multi MERCI trials. Stroke. 2009; 40:516-22.
PubMed
 
Nogueira RG, Yoo AJ, Buonanno FS, Hirsch JA.  Endovascular approaches to acute stroke, part 2: a comprehensive review of studies and trials. AJNR Am J Neuroradiol. 2009; 30:859-75.
PubMed
 
Meyers PM, Schumacher HC, Alexander MJ, Derdeyn CP, Furlan AJ, Higashida RT, et al. American Academy of Neurology.  Performance and training standards for endovascular ischemic stroke treatment. J Neurosurg. 2010; 113:149-52.
PubMed
 
Connors JJ 3rd, Sacks D, Black CM, McIff EB, Stallmeyer MJ, Cole JW, et al. Society of Interventional Radiology.  Training guidelines for intra-arterial catheter-directed treatment of acute ischemic stroke: a statement from a special writing group of the Society of Interventional Radiology. J Vasc Interv Radiol. 2009; 20:1507-22.
PubMed
 
Khatri P, Abruzzo T, Yeatts SD, Nichols C, Broderick JP, Tomsick TA, IMS I and II Investigators.  Good clinical outcome after ischemic stroke with successful revascularization is time-dependent. Neurology. 2009; 73:1066-72.
PubMed
 
Smith WS, Lev MH, English JD, Camargo EC, Chou M, Johnston SC. et al.  Significance of large vessel intracranial occlusion causing acute ischemic stroke and TIA. Stroke. 2009; 40:3834-40.
PubMed
 
Josephson SA, Saver JL, Smith WS, Merci and Multi Merci Investigators.  Comparison of mechanical embolectomy and intraarterial thrombolysis in acute ischemic stroke within the MCA: MERCI and Multi MERCI compared to PROACT II. Neurocrit Care. 2009; 10:43-9.
PubMed
 
Alexandrov AV, Grotta JC.  Arterial reocclusion in stroke patients treated with intravenous tissue plasminogen activator. Neurology. 2002; 59:862-7.
PubMed
 
Gralla J, Schroth G, Remonda L, Nedeltchev K, Slotboom J, Brekenfeld C.  Mechanical thrombectomy for acute ischemic stroke: thrombus-device interaction, efficiency, and complications in vivo. Stroke. 2006; 37:3019-24.
PubMed
 
Khatri P, Hill MD, Palesch YY, Spilker J, Jauch EC, Carrozzella JA, et al. Interventional Management of Stroke III Investigators.  Methodology of the Interventional Management of Stroke III Trial. Int J Stroke. 2008; 3:130-7.
PubMed
 
Mauldin PD, Simpson KN, Palesch YY, Spilker JS, Hill MD, Khatri P, et al. Interventional Management of Stroke III Investigators.  Design of the economic evaluation for the Interventional Management of Stroke (III) trial. Int J Stroke. 2008; 3:138-44.
PubMed
 

Letters

NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s "Cited By" API will populate this tab (http://www.crossref.org/citedby.html).

Comments

Submit a Comment
Submit a Comment

Summary for Patients

Clinical Slide Sets

Terms of Use

The In the Clinic® slide sets are owned and copyrighted by the American College of Physicians (ACP). All text, graphics, trademarks, and other intellectual property incorporated into the slide sets remain the sole and exclusive property of the ACP. The slide sets may be used only by the person who downloads or purchases them and only for the purpose of presenting them during not-for-profit educational activities. Users may incorporate the entire slide set or selected individual slides into their own teaching presentations but may not alter the content of the slides in any way or remove the ACP copyright notice. Users may make print copies for use as hand-outs for the audience the user is personally addressing but may not otherwise reproduce or distribute the slides by any means or media, including but not limited to sending them as e-mail attachments, posting them on Internet or Intranet sites, publishing them in meeting proceedings, or making them available for sale or distribution in any unauthorized form, without the express written permission of the ACP. Unauthorized use of the In the Clinic slide sets will constitute copyright infringement.

Toolkit

Want to Subscribe?

Learn more about subscription options

Advertisement
Related Articles
Related Point of Care
Topic Collections
PubMed Articles
Forgot your password?
Enter your username and email address. We'll send you a reminder to the email address on record.
(Required)
(Required)