Brief Communication: Management of Implantable Cardioverter-Defibrillators in Hospice: A Nationwide Survey FREE

Nathan Goldstein, MD; Melissa Carlson, MBA, PhD; Elayne Livote, MPH, MS, MA; and Jean S. Kutner, MD, MSPH
[+] Article and Author Information

From Mount Sinai School of Medicine, New York, New York; James J. Peters Veterans Affairs Medical Center, Bronx, New York; and University of Colorado Denver School of Medicine, Aurora, Colorado.


Portions of this study were presented at the Annual Assembly of the American Academy of Hospice and Palliative Medicine, Austin, Texas, 25–28 March 2009.

Disclaimer: The findings and interpretation reported in this manuscript do not reflect the opinions of the National Hospice and Palliative Care Organization, the Population-Based Palliative Care Research Network, the National Institute of Aging, or the National Institute of Nursing Research.

Acknowledgment: The authors thank the hospices that participated in our survey for sharing their information and policies, the National Hospice and Palliative Care Organization for sharing its list of hospices, and the Population-Based Palliative Care Research Network for assistance with the survey design.

Grant Support: By the National Institute of Aging, grants K23 AG025933 (Dr. Goldstein) and K07AG030337-01A2 (Dr. Kutner), and the National Institute of Nursing Research, grant 1K99NR010495-01 (Dr. Carlson).

Potential Conflicts of Interest: Disclosures can be viewed at www.acponline.org/authors/icmje/ConflictOfInterestForms.do?msNum=M09-1119.

Reproducible Research Statement:Study protocol, statistical code, and data set: Available from Dr. Goldstein (e-mail, Nathan.Goldstein@mssm.edu).

Requests for Single Reprints: Nathan Goldstein, MD, Department of Geriatrics, Mount Sinai School of Medicine, Box 1070, One Gustave Levy Place, New York, NY 10029; e-mail, Nathan.Goldstein@mssm.edu.

Current Author Addresses: Drs. Goldstein and Carlson: Department of Geriatrics, Mount Sinai School of Medicine, Box 1070, New York, NY 10029.

Ms. Livote: James J. Peters Veterans Affairs Medical Center, GRECC, 130 West Kingsbridge Road, Bronx, NY 10468-9938.

Dr. Kutner: Division of General Internal Medicine, B180, University of Colorado Denver School of Medicine, Academic Office 1, Room 8419, 12631 East 17th Avenue, Aurora, CO 80045.

Author Contributions: Conception and design: N. Goldstein, J.S. Kutner.

Analysis and interpretation of the data: N. Goldstein, M. Carlson, E. Livote.

Drafting of the article: N. Goldstein, E. Livote.

Critical revision of the article for important intellectual content: N. Goldstein, M. Carlson, J.S. Kutner.

Final approval of the article: N. Goldstein, M. Carlson, J.S. Kutner.

Statistical expertise: M. Carlson, E. Livote.

Obtaining of funding: N. Goldstein.

Collection and assembly of data: N. Goldstein.


Ann Intern Med. 2010;152(5):296-299. doi:10.7326/0003-4819-152-5-201003020-00007
Text Size: A A A

Background: Communication about the deactivation of implantable cardioverter-defibrillators (ICDs) in patients near the end of life is rare.

Objective: To determine whether hospices are admitting patients with ICDs, whether such patients are receiving shocks, and how hospices manage ICDs.

Design: Cross-sectional survey.

Setting: Randomly selected hospice facilities.

Participants: 900 hospices, 414 of which responded fully.

Measurements: Frequency of admission of patients with ICDs, frequency with which patients received shocks, existence of ICD deactivation policies, and frequency of deactivation.

Results: 97% of hospices admitted patients with ICDs, and 58% reported that in the past year, a patient had been shocked. Only 10% of hospices had a policy that addressed deactivation. On average, 42% (95% CI, 37% to 48%) of patients with ICDs had the shocking function deactivated.

Limitation: The study relied on the knowledge of hospice administrators.

Conclusion: Hospices are admitting patients with ICDs, and patients are being shocked at the end of life. Ensuring that hospices have policies in place to address deactivation may improve the care for patients with these devices. The authors provide a sample deactivation policy.

Primary Funding Source: National Institute of Aging and National Institute of Nursing Research.

Editors' Notes
Context

  • Patients with implantable cardioverter-defibrillators (ICDs) are being admitted to hospice care, and little is known about how these patients are managed.

Contribution

  • The investigators surveyed a nationally representative sample of 900 hospices, of which 414 responded fully. Of these, 97% admitted patients with ICDs and 58% reported that at least 1 patient had been shocked in the past year. Forty-two percent of patients with ICDs had the shocking function deactivated.

Caution

  • The study relied on the knowledge of hospice administrators.

Implication

  • Hospices should develop policies about ICD deactivation to avoid shocks at the end of life.

—The Editors

The indications for implantable cardioverter-defibrillators (ICDs) have recently been expanded (12). Because shocks from an ICD are painful and anxiety-provoking (34), some patients may choose to have the shocking function deactivated near the end of life. Conversations about deactivation are complicated and rare (56), however, and patients may not understand the role the device plays in their end-of-life care (7).

Hospices have long been leaders in caring for patients with advanced disease. Patients admitted to hospice must have a life expectancy of 6 months or less and agree to forgo life-prolonging treatment. A discussion about ICD management is therefore appropriate for a patient with an ICD who is receiving hospice care. However, no systematic examination of management for patients receiving hospice care has been performed.

Our objectives were to determine the frequency with which hospices are admitting patients with ICDs, the frequency with which such patients are receiving shocks, and the processes that hospices use to care for patients with ICDs.

We received a list of 3750 hospices from the National Hospice and Palliative Care Organization, whose membership includes more than 80% of U.S. hospices (8). From this list, we generated a stratified random sample of 100 hospices from each of the 9 U.S. Census regions. To account for the unequal probability of hospice selection by region, we included a sampling weight proportional to the inverse of the probability of a hospice being chosen by using SAS, version 9.1.3 (SAS Institute, Cary, North Carolina).

We created a novel survey instrument based on our previous work (67). The survey asked how many patients with active ICDs had been admitted and gave respondents a range of choices. It then asked whether any of the patients had been shocked in the past year (yes or no) and whether any had received multiple shocks (yes or no). Next, the survey asked whether the hospice intake forms had an item to identify patients with ICDs (yes or no), whether the hospice had a formal deactivation policy (yes or no), what percentage of patients had their device deactivated while receiving hospice care (open-ended), and whether the hospice had a strong magnet on hand for emergency deactivation (yes or no). (Questions about ICD deactivation were clearly stated as referring only to the shocking function.) We asked hospices with formal policies to submit a copy of their policy.

We then sent the survey to the Scientific Advisory Committee of the Population-Based Palliative Care Research Network (www.uchsc.edu/popcrn). The Advisory Committee provided feedback on survey readability, ease of use, and validity. We then pilot-tested the survey with several clinicians. Because the study did not collect patient data, it was exempt from review by Mount Sinai School of Medicine's institutional review board.

We mailed the survey to all selected hospices. Respondents could either fax back the paper version or complete an identical version online. We used a series of incentives, which included a pen and a $2 bill. We called nonresponding hospices to encourage participation.

We present results relating to response rate and analysis of deactivation policies as actual (unweighted) numbers and all other results as weighted percentages. We used the chi-square and t tests for bivariate analyses. When hospices answered “don't know” to a yes-or-no question, we performed a sensitivity analysis in which we assigned these responses as first “no” and then “yes” to determine the potential range of possible responses.

We reviewed the submitted ICD policies (interrater reliability, 96%) for 3 generally accepted necessary criteria (910): a prompt to identify patients with ICDs, discussion of the benefits and burdens of the device as they relate to the patient's illness, and instructions on how to reprogram the device. In addition, we examined the policies to determine whether they discussed the ethical basis for deactivation, what to do if a patient could not travel, and a process that outlined the use of a magnet in an emergency setting. These last criteria have been identified as being important (although not essential) elements of a deactivation policy (10).

Role of the Funding Source

This study was funded by the National Institute of Aging and the National Institute of Nursing Research. The funding sources had no role in the study design, data collection, or manuscript preparation.

After accounting for the 41 hospices for which we had incorrect contact information and the 24 that no longer operated as a hospice, we reduced the denominator to 835. We treated the 62 hospices that indicated that they did not admit patients with ICDs but did not provide further information as refusals, as described elsewhere (11). We received completed surveys from 414 hospices, a 50% response rate. We found no evidence that response rates differed by census region (chi-square, 11.2; P = 0.20). The Table shows the hospices' characteristics.

Table Jump PlaceholderTable.  Characteristics of Hospices and Management Practices Relating to ICDs

Almost all hospices (97%) admitted patients with active ICDs. In the past year, 76% of hospices admitted between 1 and 10 patients with an active ICD. Fifty-eight percent of hospices reported that at least 1 person was shocked in the last year, and 40% of those reported that at least 1 patient had received multiple shocks during a single episode.

Twenty percent of hospices had a question on their intake forms to identify patients with ICDs, and 10% had a deactivation policy. Hospices with a question on their intake forms were more likely to have a deactivation policy (odds ratio, 4.6 [95% CI, 2.3 to 9.2]). Twenty-five percent had a strong magnet available to deactivate an ICD; of these, 64% provided training in its use.

Of patients with active devices, an average of 42% (CI, 37% to 48%) had the shocking function deactivated while receiving hospice care. Hospices with a deactivation policy had a higher mean percentage of patients who had their devices deactivated than hospices with no such policy (73% vs. 38%; P < 0.001).

Fifteen hospices provided copies of their deactivation policies. In a content analysis, 11 had a section that prompted device identification, 15 discussed informed consent in relation to deactivation, and 14 outlined the steps to reprogram the device. In addition, 9 discussed the ethical basis for deactivation, 9 addressed what should be done if the patient could not travel, and 7 addressed emergency use of a magnet. No policy required deactivation. We synthesized elements of submitted policies to create a sample policy that contains all of the essential elements (Appendix).

We found that hospices were admitting patients with active ICDs and that patients with these devices were being shocked near the end of life. Fewer than half of patients had their devices deactivated while receiving hospice care, and a hospice policy that addressed deactivation was associated with a higher percentage of device deactivations.

Improvements in models of care for patients with ICDs and advanced disease are needed. Having a question on the hospice intake form is strongly related to having a deactivation policy, but most facilities have neither. Our results do not establish a causal relationship between having a policy and having more devices deactivated, but the relationship is intriguing, and further study is needed. By providing the sample policy, we hope to facilitate quality improvement for hospices. Our objective with the sample policy is for hospices to adapt it to their own circumstances to improve the quality of care for patients and their families.

Our sample policy is a combination of the best elements of the policies we received. We caution people who use the sample policy to remember that ICDs are complex devices that may be multifunctional in many cases. A clear understanding of these devices is important to ensure the highest-quality conversations with patients and their families. Hospices must create relationships with local electrophysiologists and representatives from device manufacturing companies to ensure that patients—especially those who cannot leave their place of residence—can have their devices reprogrammed. Willingness of nonhospice clinicians or manufacturer representatives to do home visits may vary; thus, the implications of device deactivation are important not only for hospices but also for any clinician who deals with a patient with an ICD.

To our knowledge, ours is the first nationwide study to examine the management of ICDs in patients receiving hospice care. Our study has limitations. First, we relied on the knowledge of hospice administrators and did not directly verify the data that the hospices provided; however, we did ask the survey recipients to work collaboratively with other team members when responding. Second, our response rate was low but similar to or better than that of other studies that examined practices in end-of-life care (1214), and response rates did not vary by region. Finally, we conducted our survey at the level of the hospice and collected no information about individual patients. Future work will need to determine how patient-level factors relate to ICD deactivation practices.

In conclusion, hospices enroll patients with active ICDs, and patients are shocked near the end of life. These data show an association between having a deactivation policy and a higher percentage of patients with deactivated ICDs. We hope that by providing a sample policy, hospices will engage in quality improvement activities to improve outcomes for patients with ICDs and their families.

Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, et al., Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) Investigators.  Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005; 352:225-37. PubMed
CrossRef
 
Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, et al., Multicenter Automatic Defibrillator Implantation Trial II Investigators.  Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002; 346:877-83. PubMed
 
Eckert M, Jones T.  How does an implantable cardioverter defibrillator (ICD) affect the lives of patients and their families? Int J Nurs Pract. 2002; 8:152-7. PubMed
 
Sears SF Jr, Conti JB.  Quality of life and psychological functioning of ICD patients. Heart. 2002; 87:488-93. PubMed
 
Goldstein NE, Lampert R, Bradley E, Lynn J, Krumholz HM.  Management of implantable cardioverter defibrillators in end-of-life care. Ann Intern Med. 2004; 141:835-8. PubMed
 
Goldstein NE, Mehta D, Teitelbaum E, Bradley EH, Morrison RS.  “It's like crossing a bridge”: complexities preventing physicians from discussing deactivation of implantable defibrillators at the end of life. J Gen Intern Med. 2008; 23:Suppl 12-6. PubMed
 
Goldstein NE, Mehta D, Siddiqui S, Teitelbaum E, Zeidman J, Singson M, et al..  “That's like an act of suicide”: patients' attitudes toward deactivation of implantable defibrillators. J Gen Intern Med. 2008; 23:Suppl 17-12. PubMed
 
National Hospice and Palliative Care Organization.  NHPCO Facts and Figures: Pediatric, Palliative, and Hospice Care in America. Alexandria, VA: National Hospice and Palliative Care Organization; 2009. Accessed atwww.nhpco.org/files/public/quality/Pediatric_Facts-Figures.pdfon 18 December 2009.
 
Kirk TW.  Deactivation of automatic implantable cardioverter-defibrillators in hospice and home care patients at the end of life. Home Healthc Nurse. 2008; 26:431-7. PubMed
 
Goldstein N, Kirk T, Luebke D, Wellman C.  Just a Heartbeat Away—Caring for People with Defibrillators [Audioconference]. Alexandria, VA: National Hospice and Palliative Care Organization; 2008.
 
The American Association for Public Opinion Research.  Standard Definitions: Final Dispositions of Case Codes and Outcome Rates for Surveys. 5th Edition. Lenexa, KS: American Assoc for Public Opinion Research; 2008.
 
Angus DC, Kelley MA, Schmitz RJ, White A, Popovich J Jr, Committee on Manpower for Pulmonary and Critical Care Societies (COMPACCS).  Caring for the critically ill patient. Current and projected workforce requirements for care of the critically ill and patients with pulmonary disease: can we meet the requirements of an aging population? JAMA. 2000; 284:2762-70. PubMed
 
Groeger JS, Guntupalli KK, Strosberg M, Halpern N, Raphaely RC, Cerra F, et al..  Descriptive analysis of critical care units in the United States: patient characteristics and intensive care unit utilization. Crit Care Med. 1993; 21:279-91. PubMed
 
Puntillo KA, Benner P, Drought T, Drew B, Stotts N, Stannard D, et al..  End-of-life issues in intensive care units: a national random survey of nurses' knowledge and beliefs. Am J Crit Care. 2001; 10:216-29. PubMed
 
Appendix: Sample ICD Deactivation Policy

This policy is a synthesis of actual policies submitted by responding hospices. It provides the core elements that should be included in a policy that addresses the management of these devices for patients receiving hospice care. It has not been pilot-tested and should be adapted by hospices after they have identified community partners who will assist with the reprogramming of patients' devices.

Ethical Rationale for Deactivation

Implantable cardioverter-defibrillators are often multifunctional devices that are programmed to meet an individual patient's cardiac needs. These devices are designed to terminate potentially life-threatening arrhythmias in patients, and one way that they do this is to deliver electrical shocks to the heart. Unlike other treatments these devices may deliver to correct arrhythmias, the patient may experience pain or discomfort when the ICD discharges. That an ICD is present does not automatically mean that it will fire as death approaches. Deactivation of the shocking function is not a requirement for admission to hospice but may be in line with the hospice care goal of preserving quality of life during the dying process. Defibrillators are subject to the same ethical and clinical considerations as any other medical treatment. Similarly, ICDs are subject to an analysis of potential benefits and burdens, and patients or their surrogates have the right to accept or decline its interventions. These should not be isolated decisions but should instead be made in the context of the patient's larger goals of care.

Identification of Device

At the time of evaluation and admission to hospice (regardless of the setting in which care is delivered), all patients or their families will be queried about the presence of a pacemaker or ICD. On physical examination, the chest wall of each patient should be checked for the presence of a cardiac device (devices are usually placed underneath the clavicle and may be visible or palpable). If a device is identified, the patient or his or her family should be asked whether the patient has the card that was provided at implantation, to aid in determining the nature of the device. If this card cannot be located, the hospice nurse should contact the patient's primary care physician or cardiologist to determine the nature of the device.

Informed Consent Discussion About Device Deactivation

After an ICD has been identified, the hospice nurse should engage in an informed consent discussion with the patient, family, or surrogate about the potential benefits and burdens of the device at this point in the patient's illness. To make a truly informed choice about whether to deactivate the shocking function, the nurse should emphasize the following points during the discussion:

  1. Leaving the defibrillation function on could potentially cause the patient to experience pain if the device delivers shocks near the end of life.

  2. Turning off the shocking function means the device will not be able to provide all of the available methods of life-saving therapy in the event of a potentially fatal heart rhythm. Leaving the shocking function active does not guarantee, however, that the heart will return to a normal pattern of beating after an arrhythmia.

  3. Turning off the ICD will not cause death.

  4. Turning off the ICD will not be painful, nor will a patient's death be more painful if it is turned off.

  5. Decisions about deactivating a pacemaker are often made separately from the decision to turn off a defibrillator, and they depend on the indication for the pacemaker and the patient's underlying intrinsic cardiac rhythm. Both are justifiable on ethical grounds, however, depending on the patient's overall goals of care. Deactivating a pacemaker may result in changes in a patient's symptoms. Consultation with a cardiologist or electrophysiologist before deactivation is often advisable to ensure that appropriate treatments are readily available if the pacemaker is deactivated.

Process for Reprogramming the ICD

If a decision has been made to deactivate the shocking function of the ICD, the hospice nurse will inform the medical director that the decision has been made to reprogram the device so it will no longer deliver shocks.

Note: Reprogramming an ICD in this manner will stop it from ever delivering shocks. Placing a magnet over the ICD will stop it from sensing the rhythm and delivering a shock, but only while the magnet is physically present over the device.

If the patient is ambulatory, the nurse will contact the patient's cardiologist or electrophysiologist to arrange for the patient to come to the office to deactivate the shocking function.

If the patient cannot leave his or her place of residence, then the hospice nurse will develop a plan with the attending physician for reprogramming the patient's ICD, which may include one of the following processes:

  • The patient's cardiologist or electrophysiologist or a member of his or her team will be contacted to come to the patient's place of residence to reprogram the device.

  • A member of the hospice team with special training in the deactivation of ICDs will arrange to borrow the equipment (similar to a small laptop computer) from the cardiologist or electrophysiologist and bring it to the patient's place of residence to reprogram the device. Equipment may be manufacturer-specific, so this team member must know which manufacturer made the patient's ICD.

  • A representative from the device manufacturer, after appropriate consultation with the hospice medical director or the patient's cardiologist or electrophysiologist, will come to the patient's place of residence to reprogram the ICD.

If the patient is not ambulatory, the hospice nurse will be present in the place of residence during the reprogramming process to provide emotional support to the patient, family, and surrogate.

Process for Deactivation of an ICD in an Emergent Scenario

If the decision is for the shocking function of the ICD to remain active, a magnet designed for cardiac devices should be left in the patient's place of residence in the event of an emergent scenario in which the patient is being repeatedly shocked. It should be explained to the family that if a patient is receiving repeated shocks from the ICD, then placing the magnet over the device will stop it from sensing the cardiac arrhythmia. The magnet will need to be taped in place because it only stops the ICD from sensing. (An ICD that does not sense will not deliver treatments.) Once the magnet is removed, the ICD will again begin sensing and may again deliver shocks.* The magnet is heavy and may not be comfortable if left in place for an extended period. If the family is not comfortable performing this procedure themselves, the magnet should still be left in the place of residence so that a hospice nurse who arrives in an emergent situation will have the necessary tools to suspend the shocking function of the device.

Postmortem Care

After a patient has died, the ICD will not deliver a shock. If a magnet has been taped to the chest, it can be removed as soon as a nurse has verified that the patient no longer has cardiac function. If the body is to be cremated, the funeral director should be notified of the presence of an ICD, because incinerating the battery can cause it to explode.

* At the time this protocol was written, one ICD on the market would be permanently reprogrammed to completely deactivate its shocking function when placed in the presence of a magnet for a brief period. If one is not sure of the exact specifications of a patient's ICD, the best practice is to keep the magnet in place.

Figures

Tables

Table Jump PlaceholderTable.  Characteristics of Hospices and Management Practices Relating to ICDs

References

Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, et al., Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) Investigators.  Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005; 352:225-37. PubMed
CrossRef
 
Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, et al., Multicenter Automatic Defibrillator Implantation Trial II Investigators.  Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002; 346:877-83. PubMed
 
Eckert M, Jones T.  How does an implantable cardioverter defibrillator (ICD) affect the lives of patients and their families? Int J Nurs Pract. 2002; 8:152-7. PubMed
 
Sears SF Jr, Conti JB.  Quality of life and psychological functioning of ICD patients. Heart. 2002; 87:488-93. PubMed
 
Goldstein NE, Lampert R, Bradley E, Lynn J, Krumholz HM.  Management of implantable cardioverter defibrillators in end-of-life care. Ann Intern Med. 2004; 141:835-8. PubMed
 
Goldstein NE, Mehta D, Teitelbaum E, Bradley EH, Morrison RS.  “It's like crossing a bridge”: complexities preventing physicians from discussing deactivation of implantable defibrillators at the end of life. J Gen Intern Med. 2008; 23:Suppl 12-6. PubMed
 
Goldstein NE, Mehta D, Siddiqui S, Teitelbaum E, Zeidman J, Singson M, et al..  “That's like an act of suicide”: patients' attitudes toward deactivation of implantable defibrillators. J Gen Intern Med. 2008; 23:Suppl 17-12. PubMed
 
National Hospice and Palliative Care Organization.  NHPCO Facts and Figures: Pediatric, Palliative, and Hospice Care in America. Alexandria, VA: National Hospice and Palliative Care Organization; 2009. Accessed atwww.nhpco.org/files/public/quality/Pediatric_Facts-Figures.pdfon 18 December 2009.
 
Kirk TW.  Deactivation of automatic implantable cardioverter-defibrillators in hospice and home care patients at the end of life. Home Healthc Nurse. 2008; 26:431-7. PubMed
 
Goldstein N, Kirk T, Luebke D, Wellman C.  Just a Heartbeat Away—Caring for People with Defibrillators [Audioconference]. Alexandria, VA: National Hospice and Palliative Care Organization; 2008.
 
The American Association for Public Opinion Research.  Standard Definitions: Final Dispositions of Case Codes and Outcome Rates for Surveys. 5th Edition. Lenexa, KS: American Assoc for Public Opinion Research; 2008.
 
Angus DC, Kelley MA, Schmitz RJ, White A, Popovich J Jr, Committee on Manpower for Pulmonary and Critical Care Societies (COMPACCS).  Caring for the critically ill patient. Current and projected workforce requirements for care of the critically ill and patients with pulmonary disease: can we meet the requirements of an aging population? JAMA. 2000; 284:2762-70. PubMed
 
Groeger JS, Guntupalli KK, Strosberg M, Halpern N, Raphaely RC, Cerra F, et al..  Descriptive analysis of critical care units in the United States: patient characteristics and intensive care unit utilization. Crit Care Med. 1993; 21:279-91. PubMed
 
Puntillo KA, Benner P, Drought T, Drew B, Stotts N, Stannard D, et al..  End-of-life issues in intensive care units: a national random survey of nurses' knowledge and beliefs. Am J Crit Care. 2001; 10:216-29. PubMed
 

Letters

NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s "Cited By" API will populate this tab (http://www.crossref.org/citedby.html).

Comments

Submit a Comment
No Title
Posted on March 6, 2010
Douglas R Cook
No Affiliation
Conflict of Interest: None Declared

After reading the McLellan editorial, I did some further research on the man who holds the Deputy Director position with the Obama White House Office of National Drug Control Policy. We knew at the start that he held a PhD and not a medical degree, so we could judge the comments from his expertise in addiction studies and not from what's medically best for the body hit with the stress burden of pain. But I am more than a little disturbed that his full title was not disclosed at the time of the editorial's first printing. Mr. McLellan's full title is "Deputy Director for DEMAND REDUCTION." I believe that puts a much different spin on his opinions when it becomes known that his very job is to reduce the amount of users of opioids nationally. I hold to this philosophy as a caregiver for my wife who suffers from unrelenting pain when opioid treatment is not available to her: Pain MEDICINE is for Pain PATIENTS, just as Insulin is for Diabetics. If there is to be any reduction in demand, make it come from any other area but the one for which it was made.

Conflict of Interest:

None declared

Complexities of Defibrillator Deactivation
Posted on March 12, 2010
Sandesh Dev
Duke University Medical Center
Conflict of Interest: None Declared

We strongly agree with the authors' conclusions that hospices may improve quality of care if they routinely identify patients with implantable defibrillators (ICDs) and to address ICD deactivation. In our own hospital-based heart failure cardiology and palliative care practices, we always address ICD deactivation prior to discharging patients to hospice care. However, given the complexity and difficulty of decision- making for families and caregivers during transitions in care, we have observed the occasional patient who is not ready to deactivate their ICD at time of hospital discharge. We appreciate that hospices do accept patients with functioning ICDs with the presumed goal of addressing deactivation after enrollment in hospice. We would like to highlight the study limitation noted by the authors that this study does not directly address patient-level preferences about ICD inactivation. Goldstein et al have reported in a small study that ICD recipients have not discussed ICD deactivation and are unaware that deactivation is an option(1). In our experience, we also have observed that most patients are not made aware of the possible future burden of the ICD at the time of device implantation. We have occasionally noticed a paradox in which patients, despite preferring palliative care or hospice, remain unwilling to undergo deactivation of their ICDs. Furthermore, it is possible that hospices nurses are appropriately querying hospice patients about their preferences for ICD deactivation, irrespective of the presence of a hospice ICD deactivation policy. Further, we want to make note that the median hospice length of stay in a large sample of Medicare heart failure patients was reported at only 17 days with nearly 1/3 of patients staying less than 8 days(2. Thus, a short enrollment in hospice at end of life may be an additional barrier to establishment of highest quality conversations between cardiac patients and hospice care providers. Finally, clinicians need to be aware that, unlike cancer patients, 19% of HF patients are discharged from hospice and nearly 1 in 4 patients survive more than 180 days after hospice enrollment. Thus, admission to hospice care is not a certain indication of terminal HF, which adds to the complexity of prognostication and decision-making with this disease.

References:

1. Bain, K. T., T. L. Maxwell, et al. (2009). "Hospice use among patients with heart failure." American Heart Journal 158(1): 118-125.

2. Goldstein, N., D. Mehta, et al. (2008). "That's Like an Act of Suicide" Patients' Attitudes Toward Deactivation of Implantable Defibrillators." Journal of General Internal Medicine 23(0): 7-12.

Conflict of Interest:

None declared

Re:Complexities of Defibrillator Deactivation
Posted on April 16, 2010
Nathan E Goldstein
Mount Sinai School of Medicine
Conflict of Interest: None Declared

We appreciate the authors' thoughtful response to our article. The goal of our work has always been to encourage patient-clinician communication about ICD deactivation in order to assure that patients and families can make informed choices that are consistent with their values and desires for care. As such, we do not advocate that all patients must have their devices deactivated and encourage physicians to respect patient autonomy when it comes to the management of the ICD even if, in the clinician's mind, the patient's decision appears to be counter-intuitive. The authors' comments about patients not always deciding to deactivate their ICD are in line with our clinical experiences and previously published data. Specifically, in our publication in the Annals of Internal Medicine several years ago, we found that only 21 (78%) of the 27 patients with an ICD who were reported to have had a conversation about device deactivation before death chose to deactivate the shocking function of the device(1). Patients often have had their defibrillator for several years, and may develop complex psychological relationships with respect to the device (2, 3). As one individual with an ICD stated in our focus group: "All I know is that it is there to help me and as long as I don't do wrong by it, it won't do wrong by shocking me (4)." As the authors' response illustrates when describing the percentage of heart failure patients who are discharged from hospice alive, the course of patients with end-stage heart disease can be particularly difficult to predict(5). The experience of heart failure patients underscores the importance of our recommendation that conversations should occur earlier in the course of a patient's illness so that the patient and their caregivers can begin discussions and determine if and when in the future the patient might want the device deactivated. These conversations will make decisions easier for the family should they find themselves in an emergency situation where the patient is being shocked repeatedly at the end of life(1). Only by enabling early and comprehensive communication about ICD management in the context of other discussions about the patient's overall goals of care can we assure the best quality of life and care for patients with advanced heart disease and their families.

References

1. Goldstein NE, Lampert R, Bradley EH, Lynn J, Krumholz HM. Management of Implantable Cardioverter Defibrillators in End-of-Life Care. Annals of Internal Medicine. 2004;141:835-838.

2. Dunbar SB. Psychosocial issues of patients with implantable cardioverter defibrillators. Am J Crit Care. 2005;14(4):294-303.

3. Sears SF, Conti J. Quality of Life and Psychological Functioning of ICD Patients. Heart. 2002;87:488-493.

4. Goldstein NE, Mehta D, Siddiqui S, et al. "That's like an act of suicide" patients' attitudes toward deactivation of implantable defibrillators. J Gen Intern Med. 2008;23 Suppl 1:7-12.

5. Lunney JR, Lynn J, Foley DJ, Lipson S, Guralnik JM. Patterns of functional decline at the end of life. JAMA. 2003;289(18):2387-92.

Conflict of Interest:

None declared

Re:Re:Complexities of Defibrillator Deactivation
Posted on May 4, 2010
James M Beattie
NHS Heart Improvement Programme; National Council for Palliative Care; British Heart Foundation, UK
Conflict of Interest: None Declared

We read with interest the article by Dr Goldstein and others describing the survey results of the experience and management of implantable cardioverter defibrillators (ICDs) by hospice organisations across the United States (1). This report is timely given the observed exponential growth in implant rates in recent years as guidelines for a range of device therapies have evolved. These issues are also relevant to the United Kingdom where hospices frequently admit patients whose devices remain active. Management dilemmas arise and often seem unanticipated. We have attempted to raise awareness of health professionals to the possible confounding influence of ICD therapy in advanced disease states (2) while the patient support group Arrhythmia Alliance have produced an information leaflet for recipients and their informal carers (www.heartrhythmcharity.org.uk/html/icds.html).

In 2008 a National End of Life Care Strategy was implemented in England. The long established Liverpool Care Pathway was proposed as one of the major elements of that strategy, facilitating a structured approach to care in the last few days of life, applicable particularly to those in hospital, care homes and hospices. The protocol for this health care instrument, currently in version #12, is founded on a central core, documenting a framework for end of life care provision, but the section on heart failure specifically addresses ICD deactivation (www.liv.ac.uk/mcpcil/liverpool-care-pathway/). It is clear that the practicalities of ICD deactivation remain challenging and demand close collaboration between cardiologists, electrophysiologists, cardiac nursing services and primary care physicians as well as hospice care providers. In our experience device industry representatives are often uncomfortable in undertaking defibrillator deactivation and such delegation may raise concerns about clinical responsibility (3).

We favor a patient-centered multidisciplinary team approach to prospective care planning which should facilitate the development of a consensus on timely practical ICD management; this to better ensure that evolving goals of care are comprehensively reviewed and continue to be fulfilled. Balancing the benefits and burdens of such complex therapy can be difficult (4) and good clinical navigation is essential. The sample ICD deactivation policy outlined by the authors in this brief communication provides a useful template to promote improved care for hospice patients with these devices. Compiled as a synthesis of the information provided by the US survey respondents, the resulting policy document is largely generic and thus readily applicable on an international basis. We welcome this contribution to the literature on an increasingly encountered clinical scenario.

James M Beattie

Consultant Cardiologist / National Clinical Lead, NHS Heart Improvement Programme, Birmingham UK

Emily Sam

Deputy Director of Policy Development, National Council for Palliative Care, London UK

Shai Davies

Head of Nursing / Healthcare Professional Services, British Heart Foundation, London UK

References

1. Goldstein N, Carlson M, Livote E, Kutner JS. Brief communication: management of implantable cardioverter-defibrillators in hospice: a nationwide survey. Ann Intern Med 2010; 152: 296-299. [PMID: 20194235]

2. Beattie JM. Implantable cardioverter defibrillators in patients who are reaching the end of life. London. 2007. British Heart Foundation. (available at www.bsh.org.uk/portals/2/icd%20leaflet.pdf).

3. Mueller PS, Jenkins SM, Bramstedt KA, Hayes DL. Deactivating implanted cardiac devices in terminally ill patients: practices and attitudes. PACE 2008;31: 560-568. [PMID: 18439169]

4. Dorian P. Effective and efficient use of implantable defibrillators: sometimes it's over when it's over. CMAJ 2009; 180: 599-600. [PMID: 19289799]

Conflict of Interest:

None declared

Submit a Comment

Summary for Patients

Clinical Slide Sets

Terms of Use

The In the Clinic® slide sets are owned and copyrighted by the American College of Physicians (ACP). All text, graphics, trademarks, and other intellectual property incorporated into the slide sets remain the sole and exclusive property of the ACP. The slide sets may be used only by the person who downloads or purchases them and only for the purpose of presenting them during not-for-profit educational activities. Users may incorporate the entire slide set or selected individual slides into their own teaching presentations but may not alter the content of the slides in any way or remove the ACP copyright notice. Users may make print copies for use as hand-outs for the audience the user is personally addressing but may not otherwise reproduce or distribute the slides by any means or media, including but not limited to sending them as e-mail attachments, posting them on Internet or Intranet sites, publishing them in meeting proceedings, or making them available for sale or distribution in any unauthorized form, without the express written permission of the ACP. Unauthorized use of the In the Clinic slide sets will constitute copyright infringement.

Toolkit

Want to Subscribe?

Learn more about subscription options

Advertisement
Related Articles
Journal Club
Topic Collections
PubMed Articles

Want to Subscribe?

Learn more about subscription options

Forgot your password?
Enter your username and email address. We'll send you a reminder to the email address on record.
(Required)
(Required)